Introduction

For the full video, please visit https://services.gehealthcare.com/gehcst

What you need

Step 1

              Air Filter               
  • Disconnect the pipeline sources from the wall.
  • These are the yellow and green hoses, the air and oxygen hoses, in the image.

Disconnect the pipeline sources from the wall.

These are the yellow and green hoses, the air and oxygen hoses, in the image.

1024

Step 2

  • Disconnect the oxygen hose from the machine.
  • This also includes the removal of the oxygen fitting.

Disconnect the oxygen hose from the machine.

This also includes the removal of the oxygen fitting.

Step 3

  • After the oxygen hose has been disconnected, the air hose will be disconnected as well.
  • The air line may have a compressor hose, like this unit, which will need to be removed before the air fitting can be removed.
  • Once this is done, you can look through both openings with a flashlight for debris. If there is debris, move onto the next step. Otherwise, you should reinstall the hoses to the machine and the wall

After the oxygen hose has been disconnected, the air hose will be disconnected as well.

The air line may have a compressor hose, like this unit, which will need to be removed before the air fitting can be removed.

Once this is done, you can look through both openings with a flashlight for debris. If there is debris, move onto the next step. Otherwise, you should reinstall the hoses to the machine and the wall

Step 4

  • Open both shut-off valves.
  • Verify that the inspiratory outlet is open to atmosphere (not plugged).
  • On the VCB screen:
  • Verify that the Air and O2 Flow Valves are energized ().
  • Set the Air DAC counts to 60,000 (fully open).
  • Set the O2 DAC counts to 60,000 (fully open).

Open both shut-off valves.

Verify that the inspiratory outlet is open to atmosphere (not plugged).

On the VCB screen:

Verify that the Air and O2 Flow Valves are energized ().

Set the Air DAC counts to 60,000 (fully open).

Set the O2 DAC counts to 60,000 (fully open).

Step 5

  • After 20 seconds:
  • Close both shut-off valves. Pulse the Air and O2 flow valves 5 times for 5 seconds.
  • Set the Air DAC counts to zero (closed).
  • Set the O2 DAC counts to zero (closed).
  • Note: Closing the shut-off valves before setting the flow DAC counts to zero, bleeds pressure from the system.

After 20 seconds:

Close both shut-off valves. Pulse the Air and O2 flow valves 5 times for 5 seconds.

Set the Air DAC counts to zero (closed).

Set the O2 DAC counts to zero (closed).

Note: Closing the shut-off valves before setting the flow DAC counts to zero, bleeds pressure from the system.

To reassemble your device, follow these instructions in reverse order.