Introduction
Please reference the parts list below for the disassembly diagram.
Please refer to page 7 of the device’s Technical Manual for source information.
What you need
Step 1
Valve System
- Remove the back cover, fixed by four screws.
Remove the back cover, fixed by four screws.
1024
Step 2
- Disconnect the tube from the manometer.
Disconnect the tube from the manometer.
Step 3
- Remove the manometer by unscrewing the two retaining nuts.
Remove the manometer by unscrewing the two retaining nuts.
Step 4
- Fit the existing manometer into the new valve assembly panel.
Fit the existing manometer into the new valve assembly panel.
Step 5
- Tighten the retaining nuts.
Tighten the retaining nuts.
Step 6
- Reconnect the manometer tube.
Reconnect the manometer tube.
Step 7
- Refit the front panel to the back cover with the four screws from step 1.
Refit the front panel to the back cover with the four screws from step 1.
Step 8
- Carry out the valve system performance test described above. It is recommended to record the lot number of the new valve assembly onto the Set-up and Maintenance Checklist.
Carry out the valve system performance test described above. It is recommended to record the lot number of the new valve assembly onto the Set-up and Maintenance Checklist.
To reassemble your device, follow these instructions in reverse order.
Cancel: I did not complete this guide.
Author
with 2 other contributors
Constantine Yap
Member since: 05/04/2020
337 Reputation
2 Guides authored
Badges:
3
Team
Cal Poly, Team S1-G6, Zimmerman Spring 2020
Member of Cal Poly, Team S1-G6, Zimmerman Spring 2020
CPSU-ZIMMERMAN-S20S1G6
4 Members
26 Guides authored