Introduction
What you need
Step 1
Manometer
- Remove the back cover, fixed by four screws.
Remove the back cover, fixed by four screws.
1024
Step 2
- Disconnect the tube from the manometer.
Disconnect the tube from the manometer.
Step 3
- Remove the manometer by unscrewing the two retaining nuts.
Remove the manometer by unscrewing the two retaining nuts.
Step 4
- Fit the new manometer into the front panel.
Fit the new manometer into the front panel.
Step 5
- Tighten the retaining nuts.
Tighten the retaining nuts.
Step 6
- Reconnect the manometer tube.
Reconnect the manometer tube.
Step 7
- Refit the front panel to the back cover with the four screws.
Refit the front panel to the back cover with the four screws.
Step 8
- Carry out the manometer performance test. It is recommended to record the lot number from the box label of the replacement manometer on the maintenance checklist.
Carry out the manometer performance test. It is recommended to record the lot number from the box label of the replacement manometer on the maintenance checklist.
To reassemble your device, follow these instructions in reverse order.
Cancel: I did not complete this guide.
Author
with 1 other contributor
Tyler Lin
Member since: 05/04/2020
325 Reputation
2 Guides authored
Badges:
5
+2 more badges
Team
Cal Poly, Team S1-G6, Zimmerman Spring 2020
Member of Cal Poly, Team S1-G6, Zimmerman Spring 2020
CPSU-ZIMMERMAN-S20S1G6
4 Members
26 Guides authored