Introduction
To begin, make sure the ventilator is turned off and unplugged.
These should be cleaned each day that the ventilator is in use (usually at the end of the day).
NOTE: Screw types and lengths are not specified in manuals
Step 1
Cleaning the Ulco Integrus PSV
- Absorber
- Loosen the knurled screws and remove the clear lid.
Absorber
Loosen the knurled screws and remove the clear lid.
1024
Step 2
- Remove the silicone shutters by gently lifting from the base, NOT from the flaps.
Remove the silicone shutters by gently lifting from the base, NOT from the flaps.
Step 3
- To release and remove the canister(s), undo the handle latch by turning sideways.
To release and remove the canister(s), undo the handle latch by turning sideways.
Step 4
- Push the release handle down to vertical until canisters are free.
Push the release handle down to vertical until canisters are free.
Step 5
- Push canister(s) down then out of the cradle. Then dispose of the used soda lime (see suppliers recommended procedure for disposal of soda lime).
- Wash all the absorber components. Do NOT use caustic cleaning fluids.
- The Canister(s) and silicone shutters can be autoclaved.
- The Absorber head can be put through a washer at 80 deg.C
- Dry thoroughly before assembly, low pressure warm air should be passed through the head by attaching a hose to the expiratory port of the absorber.
Push canister(s) down then out of the cradle. Then dispose of the used soda lime (see suppliers recommended procedure for disposal of soda lime).
Wash all the absorber components. Do NOT use caustic cleaning fluids.
The Canister(s) and silicone shutters can be autoclaved.
The Absorber head can be put through a washer at 80 deg.C
Dry thoroughly before assembly, low pressure warm air should be passed through the head by attaching a hose to the expiratory port of the absorber.
Step 6
- Bellows Assembly
- Loosen the four knurled screws (1) and remove the bellows canister.
Bellows Assembly
Loosen the four knurled screws (1) and remove the bellows canister.
Step 7
- Remove the bellows assembly by rotating counter-clockwise. The bellows canister may be washed or autoclaved.
Remove the bellows assembly by rotating counter-clockwise. The bellows canister may be washed or autoclaved.
Step 8
- The base disk (mushroom) should be removed from the bellows assembly before washing. Pull the latex-free rubber bag from the delrin base disk. The bellows may be washed or autoclaved. The base component (mushroom) should not be autoclaved but can be put through a washer at 80°C.
- Dry all components thoroughly before re-assembly.
- Low pressure warm air should be passed through the absorber head by attaching a hose to the scavenge port
The base disk (mushroom) should be removed from the bellows assembly before washing. Pull the latex-free rubber bag from the delrin base disk. The bellows may be washed or autoclaved. The base component (mushroom) should not be autoclaved but can be put through a washer at 80°C.
Dry all components thoroughly before re-assembly.
Low pressure warm air should be passed through the absorber head by attaching a hose to the scavenge port
To reassemble your device, follow these instructions in reverse order.
Cancel: I did not complete this guide.
One other person completed this guide.
Author
with 2 other contributors
Jacob Henriksen
Member since: 04/15/2019
232 Reputation
1 Guide authored
Badges:
8
+5 more badges
Team
Cal Poly, Team S7-G13, Paton Spring 2020
Member of Cal Poly, Team S7-G13, Paton Spring 2020
CPSU-PATON-S20S7G13
3 Members
8 Guides authored