Difficulty
Difficult
Steps
26
Time Required
35 minutes
Sections
6
- Cardinal Health Vela Ventilator Power Cable Removal
- 2 steps
- Cardinal Health Vela Ventilator Top Cover Disassembly
- 3 steps
- Cardinal Health Vela Ventilator Battery Tray Removal
- 3 steps
- Left Panel
- 4 steps
- Blender Assembly P/N 16358A
- 9 steps
- Turbine/Muffler Assembly P/N 16349
- 5 steps
Flags
2
In Progress
This guide is a work in progress. Reload periodically to see the latest changes!
Needs More Images
A few more images would make this guide’s procedures crystal clear.
BackCardinal Health Vela Ventilator
Full Screen
Options
History
Save to Favorites
Download PDF
Edit
Translate
Get Shareable Link
Embed This Guide
Notify Me of Changes
Stop Notifications
Introduction
What you need
Step 1
Cardinal Health Vela Ventilator Power Cable Removal
- Use a screwdriver to remove the two Phillips pan-head screws at the top of the plug guard.
Use a screwdriver to remove the two Phillips pan-head screws at the top of the plug guard.
1024
Step 2
- Remove the plug guard and unplug the ventilator.
Remove the plug guard and unplug the ventilator.
Step 3
Cardinal Health Vela Ventilator Top Cover Disassembly
- Remove the four Phillips #0 screws on the back side of the ventilator.
Remove the four Phillips #0 screws on the back side of the ventilator.
Step 4
- Remove all six small side panel screws along the bottom of each side of the top cover with the same screwdriver.
Remove all six small side panel screws along the bottom of each side of the top cover with the same screwdriver.
Step 5
- Slide the top cover towards the back side of the ventilator to remove it.
Slide the top cover towards the back side of the ventilator to remove it.
Step 6
Cardinal Health Vela Ventilator Battery Tray Removal
- Use your Phillips screwdriver to unscrew the 4 Phillips pan head screws positioned on each side of the battery tray.
Use your Phillips screwdriver to unscrew the 4 Phillips pan head screws positioned on each side of the battery tray.
Step 7
- Lift the battery tray out of the ventilator.
- Caution: The wires connecting the batteries to the ventilator will still be connected so lift the battery tray carefully.
Lift the battery tray out of the ventilator.
Caution: The wires connecting the batteries to the ventilator will still be connected so lift the battery tray carefully.
Step 8
- Disconnect the batteries from the ventilator by pulling apart the Battery Connector.
Disconnect the batteries from the ventilator by pulling apart the Battery Connector.
Step 9
Left Panel
- Remove the two Phillips countersink at the rear of the panel.
Remove the two Phillips countersink at the rear of the panel.
Step 10
- Remove the Phillips countersink in the upper front of the panel.
Remove the Phillips countersink in the upper front of the panel.
Step 11
- The panel is now free and can be lifted from the device.
The panel is now free and can be lifted from the device.
Step 12
- To install the left panel follow these steps in reverse order.
To install the left panel follow these steps in reverse order.
Step 13
Blender Assembly P/N 16358A
- If the rear panel is installed, remove the high and low pressure oxygen fittings from the rear panel using your 3/4" wrench. NOTE: the high pressure fitting is located above the low pressure fitting
If the rear panel is installed, remove the high and low pressure oxygen fittings from the rear panel using your 3/4" wrench. NOTE: the high pressure fitting is located above the low pressure fitting
Step 14
- If the rear panel is installed, disconnect the 40-pin ribbon cable from the SVGA out PCB.
If the rear panel is installed, disconnect the 40-pin ribbon cable from the SVGA out PCB.
Step 15
- If the front panel is installed, disconnect the green oxygen nebulizer tube.
If the front panel is installed, disconnect the green oxygen nebulizer tube.
Step 16
- Disconnect the:
- -J301, 12-Pin Connector to J301 on the main PCB
- From the blender PCB
Disconnect the:
-J301, 12-Pin Connector to J301 on the main PCB
From the blender PCB
Step 17
- Remove the bottom strip ambient air inlet filter by pinching and pulling it out.
Remove the bottom strip ambient air inlet filter by pinching and pulling it out.
Step 18
- Remove the three Phillips pan-head screws on the right side of the rear panel.
Remove the three Phillips pan-head screws on the right side of the rear panel.
Step 19
- The assembly should now be free to remove. As you remove the unit use your needle-nose pliers to disconnect the oxygen diffuser tube as you remove the blender assembly.
The assembly should now be free to remove. As you remove the unit use your needle-nose pliers to disconnect the oxygen diffuser tube as you remove the blender assembly.
Step 20
- If the exhalation valve assembly is installed, disconnect:
- J3 [2-pin (blue connection)]
- To the oxygen inlet solenoid on the blender assembly.
- J4 [2-pin (red connection)]
- To the Nebulizer solenoid assembly on upper rear of blender.
If the exhalation valve assembly is installed, disconnect:
J3 [2-pin (blue connection)]
To the oxygen inlet solenoid on the blender assembly.
J4 [2-pin (red connection)]
To the Nebulizer solenoid assembly on upper rear of blender.
Step 21
- For Instillation follow steps in reverse order.
For Instillation follow steps in reverse order.
Step 22
Turbine/Muffler Assembly P/N 16349
- Disconnect the main wire harness connector at P2 on the Turbine Driver PCBA
Disconnect the main wire harness connector at P2 on the Turbine Driver PCBA
Step 23
- Remove the 5 Phillips pan-head mounting screws.
Remove the 5 Phillips pan-head mounting screws.
Step 24
- Disconnect the gray silicone intake and output elbows by gently pulling them out from the muffler tubes.
Disconnect the gray silicone intake and output elbows by gently pulling them out from the muffler tubes.
Step 25
- The Turbine/Muffler Assembly should now be free to lift off the manifold base.
- Gently pull up on the elbows to remove
The Turbine/Muffler Assembly should now be free to lift off the manifold base.
Gently pull up on the elbows to remove
Step 26
- To Install P/N 16350 “Turbine/Muffler Assembly” repeat the steps in reverse order.
To Install P/N 16350 “Turbine/Muffler Assembly” repeat the steps in reverse order.
To reassemble your device, follow these instructions in reverse order.