Difficulty
Moderate
Steps
12
Time Required
15 minutes
Sections
4
- Cardinal Health Vela Ventilator Power Cable Removal
- 2 steps
- Cardinal Health Vela Ventilator Top Cover Disassembly
- 3 steps
- Cardinal Health Vela Ventilator Battery Tray Removal
- 3 steps
- Left Panel
- 4 steps
Flags
2
In Progress
This guide is a work in progress. Reload periodically to see the latest changes!
Needs More Images
A few more images would make this guide’s procedures crystal clear.
BackCardinal Health Vela Ventilator
Full Screen
Options
History
Save to Favorites
Download PDF
Edit
Translate
Get Shareable Link
Embed This Guide
Notify Me of Changes
Stop Notifications
Introduction
What you need
Step 1
Cardinal Health Vela Ventilator Power Cable Removal
- Use a screwdriver to remove the two Phillips pan-head screws at the top of the plug guard.
Use a screwdriver to remove the two Phillips pan-head screws at the top of the plug guard.
1024
Step 2
- Remove the plug guard and unplug the ventilator.
Remove the plug guard and unplug the ventilator.
Step 3
Cardinal Health Vela Ventilator Top Cover Disassembly
- Remove the four Phillips #0 screws on the back side of the ventilator.
Remove the four Phillips #0 screws on the back side of the ventilator.
Step 4
- Remove all six small side panel screws along the bottom of each side of the top cover with the same screwdriver.
Remove all six small side panel screws along the bottom of each side of the top cover with the same screwdriver.
Step 5
- Slide the top cover towards the back side of the ventilator to remove it.
Slide the top cover towards the back side of the ventilator to remove it.
Step 6
Cardinal Health Vela Ventilator Battery Tray Removal
- Use your Phillips screwdriver to unscrew the 4 Phillips pan head screws positioned on each side of the battery tray.
Use your Phillips screwdriver to unscrew the 4 Phillips pan head screws positioned on each side of the battery tray.
Step 7
- Lift the battery tray out of the ventilator.
- Caution: The wires connecting the batteries to the ventilator will still be connected so lift the battery tray carefully.
Lift the battery tray out of the ventilator.
Caution: The wires connecting the batteries to the ventilator will still be connected so lift the battery tray carefully.
Step 8
- Disconnect the batteries from the ventilator by pulling apart the Battery Connector.
Disconnect the batteries from the ventilator by pulling apart the Battery Connector.
Step 9
Left Panel
- Remove the two Phillips countersink at the rear of the panel.
Remove the two Phillips countersink at the rear of the panel.
Step 10
- Remove the Phillips countersink in the upper front of the panel.
Remove the Phillips countersink in the upper front of the panel.
Step 11
- The panel is now free and can be lifted from the device.
The panel is now free and can be lifted from the device.
Step 12
- To install the left panel follow these steps in reverse order.
To install the left panel follow these steps in reverse order.
To reassemble your device, follow these instructions in reverse order.